HOBO U20L water level data logger, side view with end cap fitted
HOBO U20L water level logger with end cap removed, showing the ceramic pressure sensor
HOBO U20L-04 water level logger held beside a stream with a depth gauge post

HOBO U20L Water Level Data Logger

Regular price£590.95 Inc VAT Price on application
£492.46 ex VAT @ 20%
/
Shipping calculated at checkout.
SKU U20L-04
  • In stock, only 3 left
  • Available to order

An accurate, low cost data logger designed to record water level fully compensated for barometric pressure in wells, streams, lakes etc.

Channel(s)2
HOBO Product SeriesU20L
MeasurementsBarometric Pressure, Pressure, Temperature, Water Level
Data Available ByUSB
Pressure (Absolute) and Water Level Measurements U20L-04 (4 m / 13 ft)
Operation Range 0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range 69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
Burst Pressure 310 kPa (45 psia) or 18 m (60 ft) depth
Water Level Accuracy* Typical error: ±0.1% FS, 0.4 cm (0.013 ft) water
Maximum error: ±0.2% FS, 0.8 cm (0.026 ft) water
Raw Pressure Accuracy** ±0.3% FS, 0.43 kPa (0.063 psi) maximum error
Resolution <0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water
Pressure Response Time (90%)*** <1 second at a stable temperature; measurement accuracy also depends on temperature response time
Pressure (Absolute) and Water Level Measurements U20L-01 (9 m / 30 ft)
Operation Range 0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range 69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F)
Burst Pressure 310 kPa (45 psia) or 18 m (60 ft) depth
Water Level Accuracy* Typical error: ±0.1% FS, 1.0 cm (0.03 ft) water
Maximum error: ±0.2% FS, 2.0 cm (0.06 ft) water
Raw Pressure Accuracy** ±0.3% FS, 0.62 kPa (0.09 psi) maximum error
Resolution <0.02 kPa, 0.21 cm (0.007 ft) water
Pressure Response Time (90%)*** <1 second at a stable temperature; measurement accuracy also depends on temperature response time
Pressure (Absolute) and Water Level Measurements U20L-02 (30 m / 100 ft)
Operation Range 0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range 69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
Burst Pressure 500 kPa (72.5 psia) or 40.8 m (134 ft) depth
Water Level Accuracy* Typical error: ±0.1% FS, 3.0 cm (0.1 ft) water
Maximum error: ±0.2% FS, 6.0 cm (0.2 ft) water
Raw Pressure Accuracy** ±0.3% FS, 1.20 kPa (0.17 psi) maximum error
Resolution <0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water
Pressure Response Time (90%)*** <1 second at a stable temperature; measurement accuracy also depends on temperature response time
Temperature Measurements (All Models)
Operation Range -20° to 50°C (-4° to 122°F)
Accuracy ±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F), see Plot A in manual
Resolution 0.10°C at 25°C (0.18°F at 77°F), see Plot A in manual
Response Time (90%) 10 minutes in water (typical)
Stability (Drift) 0.1°C (0.18°F) per year
Logger
Real-time Clock ±1 minute per month 0° to 50°C (32° to 122°F)
Battery 2/3 AA, 3.6 Volt lithium, factory-replaceable
Battery Life (Typical Use) 5 years with 1 minute or greater logging interval
Memory (Non-volatile) 64K bytes memory (approx. 21,700 pressure and temperature samples)
Weight Approximately 154 g (5.43 oz) in air Approximately 53.9 g (1.9 oz) in fresh water
Dimensions 3.18 cm (1.25 inches) diameter, 15.24 cm (6.0 inches) length; mounting hole 6.3 mm (0.25 inches) diameter
Wetted Materials Polypropylene housing and lanyard; Viton and Buna-N O-rings; ceramic sensor in acetyl end cap; stainless steel screws suitable for saltwater
Logging Interval Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOware User’s Guide for details.
Launch Modes Immediate start and delayed start
Offload Modes Offload while logging; stop and offload
Battery Indication Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile.
Environmental Rating: IP68
CE Compliant The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).

* Water Level Accuracy: With accurate reference water level measurement, known water density, accurate Barometric Compensation Assistant data, and a stable temperature environment.

** Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.

*** Changes in Temperature: Allow 20 minutes in water to achieve full temperature compensation of the pressure sensor. Maximum error due to rapid thermal changes is approximately 0.5%.

FAQs

How does the U20L turn pressure into water depth?
The U20L is non-vented, so it records absolute pressure: the weight of the water above the sensor plus the air pressure above the water. HOBOware Pro's Barometric Compensation Assistant subtracts the barometric pressure for the same period and converts what is left into water level. It uses the water density you choose (fresh, salt or a custom value) and a reference water level that you measure by hand at the start of the deployment.

Why does barometric compensation matter?
Air pressure changes with the weather. A change of 1 hPa (1 mbar) is roughly the same as 1 cm of water, so without compensation a passing weather front can look like the water level rising or falling.

When do I need a barometric reference logger?
For accurate water level data you need barometric pressure data for the site. The simplest option is a second U20L deployed in air rather than in the water. For example, hang it in the same well about 60 cm below ground level, but always above the water surface. Barometric pressure is fairly consistent across a region, so one reference logger within about 15 km (10 miles) can usually compensate every water level logger in that area, except during fast-moving weather. Data from a HOBO weather station can also be used.

Tip: take manual reference water level readings at the start and end of each deployment.

  • HOBO U20L Water Level Data Logger (depth range as ordered)
Rated depth

This item is priced on application. Tell us what you need and we'll send you a quote.

Need a quick answer? Use live chat. For quotes or detailed requirements, send us the details and we'll reply by email.

Or call us on 01243 558270

BUYING IN BULK, OR JUST WANT SOME HELPFUL ADVICE?

Call +44 (0)1243 558270, or fill out our contact form.

The HOBO U20L is a low-cost, research-grade water level data logger for continuously measuring water level and temperature in a wide range of underwater environments. It features 0.1% measurement accuracy, a polypropylene housing for use in both fresh and salt water, and a non-vented design for convenient and hassle-free deployment.

Key Features

  • Ideal for use in both fresh and saltwater environments, including wells, streams, lakes, wetlands, and tidal areas
  • Available for depths up to 4 m (13 ft), 9 m (30 ft) or 30 m (100 ft)
  • Durable ceramic pressure sensor withstands freezing
    HOBOware Pro software provides easy conversion to accurate water level reading, fully compensated for barometric pressure, temperature, and water density in wells, streams, lakes etc
  • Can also be used in air to record barometric pressure
  • The non-vented design eliminates installation/maintenance issues and errors associated with vent tubes, desiccant maintenance, water ingress etc.

Measures Absolute Pressure, Barometric Pressure, Temperature and Water Level

IMPORTANT Requires HOBOware Pro software and either a U-DTW-1 Waterproof Shuttle or the Base-U-4 Optic USB Base Station for configuration and data offload. See Connectivity Options above.