InTemp CX402 Bluetooth Pharmaceutical Fridge Temperature Logger (with Glycol Bottle)
InTemp CX400 Bluetooth Pharmaceutical Fridge Temperature Logger (with Glycol Bottle)
InTemp CX402 Pharmaceutical/Vaccine Storage Temperature Logger (with Glycol Bottle)
InTemp CX402 Pharmaceutical/Vaccine Storage Temperature Logger (with Glycol Bottle)
InTemp CX400 Bluetooth Pharmaceutical Fridge Temperature Logger (with Glycol Bottle)

InTemp CX402 Pharmaceutical/Vaccine Storage Temperature Logger (with Glycol Bottle)

Regular price£242.95 Inc VAT Price on application
£202.46 ex VAT @ 20%
/
Shipping calculated at checkout.
SKU CX402-T205
  • In stock, only 2 left
  • Available to order

Multi-use Bluetooth digital data logger with LCD and probe in glycol bottle
GDP/GMP compliant temperature data logger ideal for Healthcare and Life Science/Pharmaceutical storage monitoring.

Ideal applications: General pharmaceutical and vaccine storage, refrigerator, freezer, chamber, and cold room storage monitoring.

Meets global Cold Chain, FDA, and EU regulatory requirements.

Channel(s)1
Sensor(s)External Probe (Please select length)
Maximum Temperature100°C
Tolerance±0.5°C from -15° to 80°C
Battery Type2 AAA
MeasurementsTemperature
Data Available ByDisplay, Online, Mobile App, WiFi, Email, Ethernet, Wireless (Bluetooth/RF)

Logger

Radio Power

1 mW (0 dBm)

Transmission Range

Approximately 30.5 m (100 ft) line-of-sight

Wireless Data Standard

Bluetooth Low Energy (Bluetooth Smart)

Logger Operating Range

-30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)

Logging Rate

1 second to 18 hours

Time Accuracy

±1 minute per month at 25°C (77°F)

Battery Type

Two AAA 1.5 V alkaline or lithium batteries, user-replaceable

Battery Life

1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life.

Memory

128 KB (84,650 measurements, maximum)

Full Memory Download Time

Approximately 60 seconds; may take longer the farther the device is from the logger

LCD

LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range

Dimensions

9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)

Weight

90.2 g (3.18 oz)

Environmental Rating

IP54

CE

The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).

External Temperature Probe

Range

-40° to 100°C (-40° to 212°F)

Accuracy

±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F)
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F)
±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F)

Resolution

0.024°C at 25°C (0.04°F at 77°F)

Drift

<0.1°C (0.18°F) per year

Response Time

With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s
With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s
With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s

Cable Length

2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable

Ambient Temperature Sensor

Range

-30° to 70°C (-22° to 158°F)

Accuracy

±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F)
±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F)

Resolution

0.024°C at 25°C (0.04°F at 77°F)

Drift

<0.1°C (0.18°F) per year

Response Time

15 minutes typical to 90% in air moving 1 m/s

This item is priced on application. Tell us what you need and we'll send you a quote.

Need a quick answer? Use live chat. For quotes or detailed requirements, send us the details and we'll reply by email.

Or call us on 01243 558270

Buying in volume, or just want some helpful advice?

Call +44 (0)1243 558270, or fill out our contact form.

InTemp CX402-Txxx (with Glycol Bottle) data loggers monitor temperatures in refrigerators, freezers and other temperature-controlled environments to comply with global GDP requirements. The loggers communicate wirelessly via Bluetooth to mobile devices.

Used in conjunction with the free InTemp app, you can easily view data, check logger status, set alarms, and create and share secure PDF reports for streamlined reporting and regulatory compliance. The CX402 features a built-in LCD display to view current and daily min and max temperatures, advanced audible and visual alarm capabilities for notification of temperature excursions, and an internal sensor for ambient temperature monitoring.

With the optional CX Internet Gateway device, you can get alert notifications via text and/or email of temperature excursions, and automatically download data from loggers within range of the device.

Models are available with a 2 or 4 metre external temperature probe attached to a Glycol bottle (if required) used to insulate the probe and delay thermal response times.

Key Features

  • Easy-to-read display with current and daily min/max temperatures
  • Audible and visual alarms for temperatures outside pre-set range
  • Streamlined daily temperature checks simplify compliance
  • Cost-effective – no subscription fees, free mobile app, and free cloud data storage
  • Receive high/low temperature alerts via text/email with the optional CX Internet Gateway device
  • Battery-powered to guard against local power loss
  • Built to FDA, USP, EU, and WHO standards
  • Integrates with mobile and cloud platforms for a 21 CFR Part 11 compliant solution
  • Ideal for monitoring vaccines
  • Spend less time on compliance and more time with patients.

Who is it For?

The InTemp CX402-Txxx data logger (with probe in glycol) is right for you if:

  • You work for an organisation that’s required to primarily follow GDP/GMP, FDA, USP, and/or EU/CE guidelines for monitoring ambient to frozen temperatures
  • Your organisation is a pharmaceutical manufacturer, third-party logistics company, Life Science/Biotech storage and transportation company, or other organisation looking to monitor temperature-sensitive goods.

Recently viewed

Resources & Downloads